Flow and pressure measurement
Flow range: 1.5 l/s (DSR medium) 16 l/s (Spirette)
Volume resolution: 0.6 ml
Accuracy: 2 %
Pressure range: -12 to 12 kPa
Dead space: 7.2 ml (DSR medium) 35 ml (Spirette)
Resistance: < 0.15 kPa/ 0.5 lpm (DSR medium)
< 0.15 kPa/ lpm (Spirette)
Sampling frequency: 200 Hz
 
Molecular Mass measurement
Range: 20..45 g/mol
Resolution: 0.01 g/mol 
Accuracy: 0.01 g/mol (at const. temperature)
 
Airway occlusion module (option)
Modes of operation: Automatic (flow triggered), manual
Response time: < 10 ms

Closing time:

select. 50 to 1500 ms
 
CO2 measurement (option)
Principle: Mainstream, self calibrating
Measurement range: 0 to 99 mmHg, 0 to 14 %, 0 to 14 kPa
Accuracy: 2 mmHg (0 to 40 mm Hg)
5 % of read. (> 40 mmHg)
10 % of read. (> 77 mmHg)
Response time: < 120 ms
 
Oxygen measurement (option)
Principle: Side stream, laser diode
Measurement range: 5 to 100 %
Resolution: 0.1 %
Accuracy: 0.5 %
Response time: 120 ms
Sampling frequency: 100 Hz
 
NO measurement (option)  
Measurement range: 0 to 5 ppm
Detection limit: 0.02 ppb (depending on sample flow)
Rise time (T90): 100 ms
Sampling rate: 10 ms
Sample flow rate: select. 50 to 300 ml/min
(depending on sample flow)
 
A/D - module (optional, for research only)
Range: 2 V
Resolution: 1 mV
Linearity: 0.02 % at 25 C
Drift: 0.5 %, 0 – 50 C
 
General  
Temperature range: 5–40 C
Humidity tolerance: 5–95 % rel. humidity (non-condensing)
Supply voltage: 100 - 240 V, 50/60 Hz
Power required: 150 VA max
Data interface: RS 232 (standard)
Data acquisition: SpiroWare with “Add on Tools”, PC program
Weight (basic module): 5 kg
   All options 80 kg (w/o PC and printer)
Dimensions (H x W x D): 550 x 100 x 400 mm (21.7 x 4 x 15.8 inch)

System requirements
PC 486 type or higher, 50 MHz or higher
Microsoft Windows 95, 98, ME and Internet Explorer (V 4.0 or higher),
CD ROM and floppy disk drive
16 MB RAM, 10 MB free disk space
SVGA or XGA graphics adapter
(Note: PC, Printer, calibration gases and zero-air supply are not part of our delivery).