ANNOx CLD 88 Y Specifications
 

HOME

INDUSTRIAL

CLD 800 Series

CLD 800 Configuration tables 

CLD 88 Y 

CLD 88 Y
Specifications

Automotive

Clean Room
Monitoring

Stack Emission
Monitoring

Nitrates
In Liquids

MEDICAL

ENVIRONMENTAL

AMMONIA

CONTACT



Measuring ranges: four freely selectable ranges from 0.005-5 ppm
Min. detectable concentration: 0.05 ppb*
Noise at zero point: 0.025 ppb*
Zero drift < 0.02 ppm/day
Linearity (in range): ± 1% full scale
Lag time < 1 sec
Rise time (0-90%) < 0.1 sec
Temperature range: 5-40°C
Humidity tolerance: 5-95% rel. h (non condensing ambient air and sample gas)
Sample flow rate: selectable from 50-300 ml/min
Dry air use for O3 generator: internally generated (no external supply gas required)
Power required: 400 VA (incl. membrane pump and ozone scrubber)
Supply voltage: 90-250 V 50/60 Hz
Interface: RS 232 (standard)
Analog output: 4-20 mA, into 600 ohms max, 0-1 V, 0-10 V
Weight: 21 kg
Dimensions: height: 133 mm
width: 455 mm (with moulding 500 mm)
depth: 540 mm
Delivery includes: CLD 88 analyser, power cable, RS 232 cable, analog signal cable, manual.
* depending on sample flow




Standard CLD 88 Y molybdenum converter.
Options CLD 88 Y p prechamber
l calibration module for zero level and span

Specifications may be subject to change without notice