| Measuring ranges: |
four freely selectable ranges from 5–5000 ppm, with option d two per channel |
| Min. detectable concentration: |
0.25 ppm* |
| Noise at zero point (1 σ) |
0.125 ppm* |
| Lag time |
< 1 sec |
| Rise time (0-90%) |
< 1 sec |
| Temperature
range: |
5-40°C |
| Humidity
tolerance: |
5-95%
rel. h
(non condensing ambient air and sample gas) |
| Quenching |
for H2O: <1.5% of meas. value for CO2: <0.3%/vol.-% CO2 |
| Sample
flow rate: |
0.1
l/min |
| Input
pressure |
externally
stabilised within ± 3 mbar
|
| Dry
air use for O3 generator: |
internally
generated
(no external supply gas required) |
| Power
required: |
400
VA (incl. membrane pump and ozone scrubber) |
| Supply
voltage: |
100-230
V 50/60 Hz |
| Interface: |
RS
232 (standard) |
| Analog
output: |
4-20
mA, into 500 ohms max, 0-1 V, 0-10 V |
| Weight: |
26
kg |
| Dimensions: |
height:
133 mm
width: 450 mm (with moulding 495 mm)
depth: 545 mm |
| Delivery
includes: |
CLD
822 M h analyser, power cable,
analog signal cable, manual. |
* depending on filter setting
| Standard |
CLD 822 M h metal converter, hot tubing |
| Options | S steel converter d dual sample gas inlet MM d dual channel NOX/NOX |
Specifications
may be subject to change without notice
|