HOME
INDUSTRIAL
CLD
700 AL
CLD
700 AL Specifications
Automotive
Clean
Room
Monitoring
Stack
Emission
Monitoring
Nitrates
In Liquids
MEDICAL
ENVIRONMENTAL AMMONIA
CONTACT
|
| Measuring
ranges: |
0-0.1,
0-10, 0-100 ppm
(analog output adjustable 0-0.05, 0-0.5, 0-5, 0-50 ppm), NO2
max. 1 ppm |
| Minimum
detectable concentration: |
0.001
ppm |
| Noise
at zero point (1 sigma): |
0.0005
ppm |
| Zero
drift: |
none |
| Linearity
(in range): |
±
1% FS |
| Lagtime: |
7 sec |
| Rise
time (0-90%): |
2-100
sec selectable (range dependent) |
| Temperature
range: |
10-40
°C |
| Humidity
tolerance: |
5-95%
rel. h
(non-condensing, ambient air and sample gas) |
| Sample
flow rate: |
0.7
l/min |
| Dry
air use (for O3 generator): |
0.35
l/min |
| Power
required: |
530
VA
(incl. membrane pump and ozone scrubber) |
| Supply
voltage: |
230
V/50 Hz, 115 V/60 Hz (± 10%) |
| Interface: |
RS
232 (standard) |
| Analogue
output: |
selectable
1 V, 10 V into 500k ohms min.
20 mA into 600 ohms max. (isolated) |
| Weight: |
23
kg |
| Dimensions: |
height:
133 mm (5 1/4"; 3 HU),
with feet 158 mm (6 1/4")
width: 483 mm (19")
depth: 588 mm (23 1/8") |
| Delivery
includes: |
Analyser
CLD 700 AL analyser, membrane vacuum pump,
ozone scrubber, power cables,
RS232 cable, tubing, manual. |
ECO PHYSICS reseves the right to change
these specifications without notice
|