Measuring ranges: four freely selectable ranges
from 5–5000 ppb
Min. detectable concentration: 0.1 ppb*
Noise at zero point: 0.05 ppb*
Lag time < 1 sec
Rise time (0-90%) < 60 sec
Temperature range: 15–30 C
Humidity tolerance: 0–95 % rel. h
(non-condensing, ambient air
and sample gas)
Sample flow rate: 1 l/min
Input pressure ambient, to be stabilized
within 3 mbar
Power required: < 600 VA
(incl. membrane pump)
Supply voltage: 100 - 230 V/50–60 Hz
Interface: RS 232 (optional LAN)
Analog output: 4–20 mA into 500 Ω max.
0–1 V; 0–10 V
Weight: 40 kg
Dimensions: height: 178 mm (7 ”)
width: 450 mm (19 ”)  with moulding: 495 mm
depth: 545 mm
Delivery includes: PAS 87 analyzer, power cable, signal cable,
operator’s manual
Standard:PAS 87
Options(NH4 + pending)

* depending on filter setting